Standards
| Model | ASI-STC360 |
|---|---|
| Workroom dimensions | 1200 × 800 × 500 |
| External dimensions | 2460 × 1300 × 1350 |
| Workroom capacity | 480 L |
| Salt water tank volume | 30 L |
| Pure water tank volume | 35 L |
| Working power |
Three Phases AC380V/50HZ (60Hz is Option) 12.5KW |
| Testing sample space | 1000 × 600 × 300 |
| Opening Method | Stainless steel air spring support |
| The temperature range of the salt spray chamber | 15°C ~ 65°C |
| The temperature range of saturated barrel | RT (room temperature) ~ 70°C |
| Relative humidity of the salt spray chamber | 30% ~ 98%RH (± 3%, adjustable, refer to the temperature and humidity corresponding chart) |
| Temperature accuracy |
Temperature uniformity: ≤ 2°C; Temperature volatility: ≤ ± 1°C |
| Salt spray deposition rate | 1 ~ 2 ml/80cm²/h |
| Test methods |
Programmable spray, moist heat, dry, standard climate |
| Test conditions |
A. Saline spray test: NSS test: 35°C ± 2°C; B. Humidity and heat test:
C. Dry test:
|
| Standard accessories |
1 set of V-type/O-type storage racks, 1 glass nozzle, 2 bottles of sodium sodium (500g/bottle), 1 plastic rust barrel (5L) |
| Air compressor Requirements | Prepared by the customer (or buy LISUN LS-EU800W2-55-LS-US800W2-55L air compressor): air inlet pressure 0.4M or above, two-stage pressure regulation, spray pressure 0.07 ~ 0.17Mp (adjustable) |
| Working environment requirements |
The ambient temperature is 5°C ~ 30°C. Ventilation is recommended. When the temperature exceeds 30°C, it is recommended to adjust the room temperature through appropriate methods. The ambient humidity is below 80%Rh. |