Technical Specifications
| Load Frame Capacity | 50 kN or higher |
| Load Frame Stiffness | 100 kN/mm or higher |
| Load Frame Overcapacity | 25% or higher |
| Load Frame Dimensions |
Maximum 1800 mm height; maximum 800 × 600 mm lateral dimensions |
| Load Frame Weight | Maximum 200 kg |
| Noise Level | Maximum 32 dB at 2 m at full crosshead movement speed |
| Compliance | CE compliant loading frame |
| Crosshead Displacement Resolution | 0.1 µm or better |
| Crosshead Displacement Accuracy | ±50 µm or better |
| Crosshead Calibration | As per ISO 9513 |
| Crosshead Speed | 0.0001 mm/min or lower to 250 mm/min or higher at 50 kN |
| Pneumatic Ports | Built-in ports for compressor connection and pneumatic grips |
| Controller Processing Rate | 160 MHz or higher |
| Computer Data Acquisition | 1000 Hz or higher |
| Computer Interface | USB 2.0 or higher |
| External Device Connections | 4 connectors for LVDTs, strain gauges, extensometers, etc. |
| External Device Compatibility | Generic external devices; not restricted to proprietary devices |
| DAQ Interface | Configurable to interface with existing branded DAQs, subject to compatibility |
| Load Cell Capacity | 50 kN |
| Load Cell Type | Shear web / pancake type |
| Load Cell Accuracy |
0.2% of applied load from 0.2% to 100% of capacity |
| Load Cell Standards | ASTM E4 and ISO 7500-1 |
| Load Cell Resolution | Better than 1 part in 8,300,000 |
| Power Requirement | 230–240 V AC, single phase, ≤5 A |