+91-11-25920433
+91-11-41428065
asipl51@gmail.com
Model: ASI- 2000 PC
SYSTEM
Microprocessor. based compression testing machine capacity of 2000 kN with pace rate controller.
Operation
Electrically –Hydraulically operated
Display
Digital Display
Capacity
2000 kN.
Certification
NABL certified and should meet the specifications of IS 14858 (2000)/ ISi marked.
Lowest load measuring Range
0.1 kN
Specimen size handling capacity
1-Cubes of 100 mm and 150 mm size
2-Cylinders of 100 mm and 150 mm in diameter with 200mm and 300mm height respectively.
3-Minimum height of cylindrical specimen being tested is 300 mm + clearance
4-Blocks of size 400 mm x 200 mm x 200 mm
Maximum vertical clearance between
370 mm
Maximum horizontal distance between side plates
340 mm
Piston stroke length
50mm
Platen size
420 mm x 220 mm
Piston Diameter
222.2 mm
Voltage
Single Phase 220 V 50 Hz, AC Supply
CVT
One No. 50 VA CVT
Operating Instruction
One copy of operating instruction
must be provided.
Tool Box
At least two normal spanner with one Y Spanner Must Be provided
DPU
Digital DPU with calibration chart
Must be provided
No of spares
Five Spacers with locating ping
Load measuring device
One digital load Indicator with pace rate indicator
Loading Rate controlling system
1-Automatic Pace rate controlled
2-Microprosser based controlled
3-Loading rate as per IS 14858 ( 2000 ). If preset speed of pacer has been chosen the pacer speed shall remain within + 5 percent of the specified speed over the operating range.
Safety
1-OSHA Standard safety guarding
2- Well maintained overload capacity
Allen Key
One 3 mm Allen Key
Compression
Hydraulic Based Compression
Hydraulic Connectivity
Appropriate connecting pipe with end coupling
Cabinet for motors and other accessories
Motorized pumping unit consisting of
pressure release valve, all must be enclose and fitted in a single cabinet.
Stress Measurement
Stress measurement setup should be automatic and displayed digitally
Data Storage / download
1- Data storage facility of 2000 records
should be present.
2- Data should be downloadable in
appropriate standard format with
suitable port
Tamping Bar
Kelley Ball Penetration Apparatus
Pumping Units